Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Tytuł pozycji:

Graphical and statistical analyses of the oculocardiac reflex during a non-invasive intracranial pressure measurement.

Tytuł:
Graphical and statistical analyses of the oculocardiac reflex during a non-invasive intracranial pressure measurement.
Autorzy:
Hamarat Y; Health Telematics Science Institute, Kaunas University of Technology, Kaunas, Lithuania.
Bartusis L; Health Telematics Science Institute, Kaunas University of Technology, Kaunas, Lithuania.
Deimantavicius M; Health Telematics Science Institute, Kaunas University of Technology, Kaunas, Lithuania.
Siaudvytyte L; Eye Clinic, Lithuanian University of Health Sciences, Kaunas, Lithuania.
Januleviciene I; Eye Clinic, Lithuanian University of Health Sciences, Kaunas, Lithuania.
Ragauskas A; Health Telematics Science Institute, Kaunas University of Technology, Kaunas, Lithuania.
Bershad EM; Department of Neurology, Baylor College of Medicine, Houston, Texas, United States of America.
Fandino J; Department of Neurosurgery, Kantonsspital Aarau, Aarau, Switzerland.
Kienzler J; Department of Neurosurgery, Kantonsspital Aarau, Aarau, Switzerland.
Remonda E; Department of Neurosurgery, Kantonsspital Aarau, Aarau, Switzerland.
Matijosaitis V; Department of Neurology, Lithuanian University of Health Sciences, Kaunas, Lithuania.
Rastenyte D; Department of Neurology, Lithuanian University of Health Sciences, Kaunas, Lithuania.
Petrikonis K; Department of Neurology, Lithuanian University of Health Sciences, Kaunas, Lithuania.
Berskiene K; Sports Institute, Lithuanian University of Health Sciences, Kaunas, Lithuania.
Zakelis R; Health Telematics Science Institute, Kaunas University of Technology, Kaunas, Lithuania.
Źródło:
PloS one [PLoS One] 2018 Apr 19; Vol. 13 (4), pp. e0196155. Date of Electronic Publication: 2018 Apr 19 (Print Publication: 2018).
Typ publikacji:
Journal Article; Research Support, Non-U.S. Gov't
Język:
English
Imprint Name(s):
Original Publication: San Francisco, CA : Public Library of Science
MeSH Terms:
Diagnostic Techniques and Procedures*/instrumentation
Intracranial Pressure*
Models, Statistical*
Reflex, Oculocardiac*
Adult ; Aged ; Data Interpretation, Statistical ; Female ; Heart Rate ; Humans ; Male ; Middle Aged ; Young Adult
References:
Br J Ophthalmol. 1998 May;82(5):589. (PMID: 9713075)
Br J Anaesth. 2008 Aug;101(2):234-8. (PMID: 18524784)
Zhonghua Yan Ke Za Zhi. 1991 Jan;27(1):34-6. (PMID: 2060403)
Can Anaesth Soc J. 1983 Jul;30(4):360-9. (PMID: 6871777)
World Neurosurg. 2016 May;89:647-653.e1. (PMID: 26724629)
Am J Ophthalmol. 1987 Jul 15;104(1):61-3. (PMID: 3300354)
Neurol Res. 2014 Jul;36(7):607-14. (PMID: 24620972)
J Clin Monit Comput. 2016 Dec;30(6):889-893. (PMID: 26438656)
ScientificWorldJournal. 2010 May 04;10:811-7. (PMID: 20454763)
Anaesthesia. 1984 Jun;39(6):524-8. (PMID: 6742383)
J Am Coll Cardiol. 1998 Mar 1;31(3):593-601. (PMID: 9502641)
Am J Ophthalmol. 1992 Sep 15;114(3):307-10. (PMID: 1524119)
ScientificWorldJournal. 2010 Jul 20;10:1416-23. (PMID: 20661534)
Acta Neurochir Suppl. 2005;95:357-61. (PMID: 16463881)
J Neurosurg Anesthesiol. 2009 Jul;21(3):187-95. (PMID: 19542994)
J Maxillofac Oral Surg. 2014 Dec;13(4):373-7. (PMID: 26224999)
Invest Ophthalmol Vis Sci. 1986 Jul;27(7):1160-4. (PMID: 3721794)
Eye (Lond). 2006 Oct;20(10):1130-44. (PMID: 17019411)
Mol Med. 2009 May-Jun;15(5-6):125-6. (PMID: 19287512)
Brain Res Brain Res Rev. 2005 Nov;49(3):555-65. (PMID: 16269319)
Neurology. 2012 May 22;78(21):1684-91. (PMID: 22573638)
Br J Ophthalmol. 1962 Feb;46(2):65-98. (PMID: 18170762)
Pacing Clin Electrophysiol. 2010 Nov;33(11):1407-17. (PMID: 20663071)
J AAPOS. 2014 Apr;18(2):151-5. (PMID: 24698612)
Acta Anaesthesiol Scand. 2007 Jul;51(6):759-62. (PMID: 17488312)
Entry Date(s):
Date Created: 20180420 Date Completed: 20180723 Latest Revision: 20240323
Update Code:
20240323
PubMed Central ID:
PMC5909620
DOI:
10.1371/journal.pone.0196155
PMID:
29672564
Czasopismo naukowe
Purpose: This study aimed to examine the incidence of the oculocardiac reflex during a non-invasive intracranial pressure measurement when gradual external pressure was applied to the orbital tissues and eye.
Methods: Patients (n = 101) and healthy volunteers (n = 56) aged 20-75 years who underwent a non-invasive intracranial pressure measurement were included in this retrospective oculocardiac reflex analysis. Prespecified thresholds greater than a 10% or 20% decrease in the heart rate from baseline were used to determine the incidence of the oculocardiac reflex.
Results: None of the subjects had a greater than 20% decrease in heart rate from baseline. Four subjects had a greater than 10% decrease in heart rate from baseline, representing 0.9% of the total pressure steps. Three of these subjects were healthy volunteers, and one was a glaucoma patient.
Conclusion: The incidence of the oculocardiac reflex during a non-invasive intracranial pressure measurement procedure was very low and not associated with any clinically relevant effects.
Zaloguj się, aby uzyskać dostęp do pełnego tekstu.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies