Multicenter Evaluation of the Quantra QPlus System in Adult Patients Undergoing Major Surgical Procedures.

Tytuł:
Multicenter Evaluation of the Quantra QPlus System in Adult Patients Undergoing Major Surgical Procedures.
Autorzy:
Groves DS; From the Department of Anesthesiology, University of Virginia, Charlottesville, Virginia.
Welsby IJ; Department of Anesthesiology, Duke University, Durham, North Carolina.
Naik BI; From the Department of Anesthesiology, University of Virginia, Charlottesville, Virginia.
Tanaka K; Department of Anesthesiology, University of Maryland, Baltimore, Maryland.
Hauck JN; Department of Anesthesiology, Duke University, Durham, North Carolina.
Greenberg CS; Department of Hematology/Oncology, Medical University of South Carolina, Charleston, South Carolina.
Winegar DA; HemoSonics, LLC, Charlottesville, Virginia, and Durham, North Carolina.
Viola F; HemoSonics, LLC, Charlottesville, Virginia, and Durham, North Carolina.
Źródło:
Anesthesia and analgesia [Anesth Analg] 2020 Apr; Vol. 130 (4), pp. 899-909.
Typ publikacji:
Journal Article; Multicenter Study; Observational Study; Research Support, N.I.H., Extramural
Język:
English
Imprint Name(s):
Publication: 1998- : Baltimore, Md. : Lippincott Williams & Wilkins
Original Publication: Cleveland, International Anesthesia Research Society.
MeSH Terms:
Blood Coagulation*
Blood Coagulation Tests/*instrumentation
Cardiac Surgical Procedures/*methods
Monitoring, Intraoperative/*instrumentation
Thrombelastography/*instrumentation
Adult ; Aged ; Aged, 80 and over ; Algorithms ; Blood Coagulation Tests/methods ; Blood Loss, Surgical ; Blood Viscosity ; Cardiopulmonary Bypass ; Elasticity ; Female ; Fibrinogen/analysis ; Humans ; Male ; Middle Aged ; Monitoring, Intraoperative/methods ; Platelet Count ; Prospective Studies ; ROC Curve ; Reproducibility of Results ; Thrombelastography/methods ; Whole Blood Coagulation Time
Grant Information:
R01 HL121232 United States HL NHLBI NIH HHS
Substance Nomenclature:
9001-32-5 (Fibrinogen)
Entry Date(s):
Date Created: 20200208 Date Completed: 20200507 Latest Revision: 20200507
Update Code:
20240105
DOI:
10.1213/ANE.0000000000004659
PMID:
32032102
Czasopismo naukowe
Background: The management of perioperative bleeding and the optimization of the available therapies are subjects of significant clinical interest. Clinical guidelines recommend the use of whole blood viscoelastic testing devices to target the utilization of blood products during major surgical procedures. The Quantra QPlus System is a new cartridge-based viscoelastic testing device based on an innovative ultrasound technology. The aim of this study was to evaluate this new system in a surgical population.
Methods: Two hundred seventy-seven adult subjects were enrolled in a multicenter, prospective observational study consisting primarily of patients undergoing cardiac and major orthopedic surgeries. Samples were obtained at multiple time points for testing on the Quantra QPlus System, the rotational thromboelastometry (ROTEM) delta, and standard coagulation tests. Quantra measurements included Clot Time (CT), Heparinase Clot Time (CTH), Clot Time Ratio (CTR), Clot Stiffness (CS), Fibrinogen (FCS), and Platelet (PCS) Contributions to CS. Data analyses included assessment of the concordance of Quantra parameters with a series of clinical composite indexes formed on the basis of standard coagulation tests in 3 domains representing increased, decreased, and normal/subclinical coagulation function. Linear regression and receiver operator characteristic (ROC) analyses of Quantra parameters with corresponding parameters from ROTEM assays were also performed.
Results: The accuracy (overall percent agreement or ratio of true positives and true negatives over the entire population) between the Quantra and the composite indexes was between 72% and 98% depending on the specific parameter. Linear regression analysis indicated that the correlation between ROTEM delta and Quantra was very strong with r values ranging between 0.84 and 0.89. Results from ROC analysis demonstrated sensitivities and specificities in the 80%-90% range when QPlus parameters were used to discriminate ROTEM threshold values currently used in goal-directed treatment algorithms.
Conclusions: This study demonstrated that the Quantra QPlus System is strongly correlated with a well-established viscoelastic testing device and its parameters effectively represent the results from multiple standard laboratory assays. The Quantra has been designed to operate at the point of care with the potential to provide rapid and comprehensive results to aid in the management of coagulopathic patients.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies