Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Tytuł pozycji:

Twin robotic x-ray system in small bone and joint trauma: impact of cone-beam computed tomography on treatment decisions.

Tytuł:
Twin robotic x-ray system in small bone and joint trauma: impact of cone-beam computed tomography on treatment decisions.
Autorzy:
Grunz JP; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany. Grunz_.
Pennig L; Institute for Diagnostic and Interventional Radiology, Faculty of Medicine and University Hospital Cologne, University of Cologne, Kerpener Straße 62, 50937, Cologne, Germany.
Fieber T; Department of Trauma, Hand, Plastic and Reconstructive Surgery, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Gietzen CH; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Heidenreich JF; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Huflage H; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Gruschwitz P; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Kuhl PJ; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Petritsch B; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Kosmala A; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Bley TA; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Gassenmaier T; Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Oberdürrbacher Straße 6, 97080, Würzburg, Germany.
Źródło:
European radiology [Eur Radiol] 2021 Jun; Vol. 31 (6), pp. 3600-3609. Date of Electronic Publication: 2020 Dec 05.
Typ publikacji:
Journal Article
Język:
English
Imprint Name(s):
Original Publication: Berlin : Springer International, c1991-
MeSH Terms:
Fractures, Bone*/diagnostic imaging
Fractures, Bone*/therapy
Robotic Surgical Procedures*
Cone-Beam Computed Tomography ; Humans ; Radiography ; X-Rays
References:
Court-Brown CM, Caesar B (2006) Epidemiology of adult fractures: a review. Injury 37:691–697. (PMID: 10.1016/j.injury.2006.04.130)
Bancroft LW, Kransdorf MJ, Adler R et al (2015) ACR Appropriateness Criteria acute trauma to the foot. J Am Coll Radiol 12:575–581. (PMID: 10.1016/j.jacr.2015.02.018)
Mosher TJ, Kransdorf MJ, Adler R et al (2015) ACR Appropriateness Criteria acute trauma to the ankle. J Am Coll Radiol 12:221–227. (PMID: 10.1016/j.jacr.2014.11.015)
Torabi M, Lenchik L, Beaman FD et al (2019) ACR Appropriateness Criteria® acute hand and wrist trauma. J Am Coll Radiol 16:S7–S17. (PMID: 10.1016/j.jacr.2019.02.029)
Ricci PM, Boldini M, Bonfante E et al (2019) Cone-beam computed tomography compared to X-ray in diagnosis of extremities bone fractures: a study of 198 cases. Eur J Radiol Open 6:119–121. (PMID: 10.1016/j.ejro.2019.01.009)
De Smet E, De Praeter G, Verstraete KLA et al (2015) Direct comparison of conventional radiography and cone-beam CT in small bone and joint trauma. Skeletal Radiol 44:1111–1117. (PMID: 10.1007/s00256-015-2127-3)
Brink M, Steenbakkers A, Holla M et al (2019) Single-shot CT after wrist trauma: impact on detection accuracy and treatment of fractures. Skeletal Radiol 48:949–957. (PMID: 10.1007/s00256-018-3097-z)
Brenner DJ, Hall EJ (2007) Computed tomography — an increasing source of radiation exposure. N Engl J Med 357:2277–2284. (PMID: 10.1056/NEJMra072149)
Vanhoenacker F, Desimpel J (2017) Cone beam CT of trauma of small bones and joints. Semin Musculoskelet Radiol 21:290–302. (PMID: 10.1055/s-0037-1606134)
Konda SR, Goch AM, Haglin J, Egol KA (2018) Ultralow-dose CT (REDUCTION protocol) for extremity fracture evaluation is as safe and effective as conventional CT. J Orthop Trauma 32:216–222. (PMID: 10.1097/BOT.0000000000001137)
Alagic Z, Bujila R, Enocson A, Srivastava S, Koskinen SK (2020) Ultra-low-dose CT for extremities in an acute setting: initial experience with 203 subjects. Skeletal Radiol 49:531–539.
Posadzy M, Desimpel J, Vanhoenacker F (2018) Cone beam CT of the musculoskeletal system: clinical applications. Insights Imaging 9:35–45. (PMID: 10.1007/s13244-017-0582-1)
Veldhoen S, Schöllchen M, Hanken H et al (2017) Performance of cone-beam computed tomography and multidetector computed tomography in diagnostic imaging of the midface: a comparative study on Phantom and cadaver head scans. Eur Radiol 27:790–800. (PMID: 10.1007/s00330-016-4387-2)
Huang AJ, Chang CY, Thomas BJ, MacMahon PJ, Palmer WE (2015) Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects. Skeletal Radiol 44:797–809.
Carrino JA, Al Muhit A, Zbijewski W et al (2014) Dedicated cone-beam CT system for extremity imaging. Radiology 270:816–824. (PMID: 10.1148/radiol.13130225)
Zbijewski W, De Jean P, Prakash P et al (2011) A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. Med Phys 38:4700–4713. (PMID: 10.1118/1.3611039)
Demehri S, Muhit A, Zbijewski W et al (2015) Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system. Eur Radiol 25:1742–1751. (PMID: 10.1007/s00330-014-3546-6)
Pallaver A, Honigmann P (2019) The role of cone-beam computed tomography (CBCT) scan for detection and follow-up of traumatic wrist pathologies. J Hand Surg Am 44:1081–1087. (PMID: 10.1016/j.jhsa.2019.07.014)
Gibney B, Smith M, Moughty A, Kavanagh EC, Hynes D, MacMahon PJ (2019) Incorporating cone-beam CT into the diagnostic algorithm for suspected radiocarpal fractures: a new standard of care? Am J Roentgenol 213:1117–1123.
Hirschmann A, Pfirrmann CWA, Klammer G, Espinosa N, Buck FM (2014) Upright cone CT of the hindfoot: comparison of the non-weight-bearing with the upright weight-bearing position. Eur Radiol 24:553–558.
Grunz J-P, Kunz AS, Gietzen CH et al (2019) 3D cone-beam CT of the ankle using a novel twin robotic X-ray system: assessment of image quality and radiation dose. Eur J Radiol 119:108659. (PMID: 10.1016/j.ejrad.2019.108659)
Grunz J-P, Gietzen CH, Kunz AS et al (2020) Twin robotic X-ray system for 3D cone-beam CT of the wrist: an evaluation of image quality and radiation dose. AJR Am J Roentgenol 214:422–427. (PMID: 10.2214/AJR.19.21911)
Falkowski AL, Kovacs BK, Schwartz FR, Benz RM, Stieltjes B, Hirschmann A (2020) Comparison of 3D X-ray tomography with computed tomography in patients with distal extremity fractures. Skeletal Radiol 49(12):1965–1975.
Hart D, Wall B (2002) Radiation exposure of the UK population from medical and dental X-ray examinations. Natl Radiol Prot Board UK NRPB-W4.
Saltybaeva N, Jafari ME, Hupfer M, Kalender WA (2014) Estimates of effective dose for CT scans of the lower extremities. Radiology 273:153–159. (PMID: 10.1148/radiol.14132903)
Massey FJ (1951) The Kolmogorov-Smirnov test for goodness of fit. J Am Stat Assoc 46:68–78. (PMID: 10.1080/01621459.1951.10500769)
Wilcoxon F (1945) Individual comparisons by ranking methods. Biom Bull 1:80. (PMID: 10.2307/3001968)
Mann HB, Whitney DR (1947) On a test of whether one of two random variables is stochastically larger than the other. Ann Math Stat 18:50–60. (PMID: 10.1214/aoms/1177730491)
McNemar Q (1947) Note on the sampling error of the difference between correlated proportions or percentages. Psychometrika 12:153–157. (PMID: 10.1007/BF02295996)
Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 20:37–46. (PMID: 10.1177/001316446002000104)
Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159. (PMID: 10.2307/2529310)
Yi JW, Park HJ, Lee SY et al (2017) Radiation dose reduction in multidetector CT in fracture evaluation. Br J Radiol 90:20170240. (PMID: 10.1259/bjr.20170240)
Faccioli N, Foti G, Barillari M, Atzei A, Mucelli RP (2010) Finger fractures imaging: accuracy of cone-beam computed tomography and multislice computed tomography. Skeletal Radiol 39:1087–1095.
Nardi C, Salerno S, Molteni R et al (2018) Radiation dose in non-dental cone beam CT applications: a systematic review. Radiol Med 123:765–777. (PMID: 10.1007/s11547-018-0910-7)
Brenner DJ, Doll R, Goodhead DT et al (2003) Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci 100:13761–13766. (PMID: 10.1073/pnas.2235592100)
Suresh P, Ninan T (2013) Computed tomography of hand and wrist. In: Davies AM, Grainger AJ, James SJ (eds) Imaging of the hand and wrist - techniques and applications. Springer, Berlin Heidelberg, pp 23–36. (PMID: 10.1007/174_2012_754)
Müller FC, Gosvig KK, Børgesen H et al (2020) Dual-energy CT for suspected radiographically negative wrist fractures: a prospective diagnostic test accuracy study. Radiology 296:596–602. (PMID: 10.1148/radiol.2020192701)
Petritsch B, Kosmala A, Weng AM et al (2017) Vertebral compression fractures: third-generation dual-energy CT for detection of bone marrow edema at visual and quantitative analyses. Radiology 284:161–168. (PMID: 10.1148/radiol.2017162165)
Tuominen EKJ, Kankare J, Koskinen SK, Mattila KT (2013) Weight-bearing CT imaging of the lower extremity. AJR Am J Roentgenol 200:146–148. (PMID: 10.2214/AJR.12.8481)
Grant Information:
Z-2/CSP-06 Interdisziplinäres Zentrum für Klinische Forschung, Universitätsklinikum Würzburg
Contributed Indexing:
Keywords: Cone-beam computed tomography; Extremities; Fractures, bone; Radiography
Entry Date(s):
Date Created: 20201206 Date Completed: 20210520 Latest Revision: 20210623
Update Code:
20240104
PubMed Central ID:
PMC8128787
DOI:
10.1007/s00330-020-07563-5
PMID:
33280057
Czasopismo naukowe
Objectives: Trauma evaluation of extremities can be challenging in conventional radiography. A multi-use x-ray system with cone-beam CT (CBCT) option facilitates ancillary 3-D imaging without repositioning. We assessed the clinical value of CBCT scans by analyzing the influence of additional findings on therapy.
Methods: Ninety-two patients underwent radiography and subsequent CBCT imaging with the twin robotic scanner (76 wrist/hand/finger and 16 ankle/foot/toe trauma scans). Reports by on-call radiologists before and after CBCT were compared regarding fracture detection, joint affliction, comminuted injuries, and diagnostic confidence. An orthopedic surgeon recommended therapy based on reported findings. Surgical reports (N = 52) and clinical follow-up (N = 85) were used as reference standard.
Results: CBCT detected more fractures (83/64 of 85), joint involvements (69/53 of 71), and multi-fragment situations (68/50 of 70) than radiography (all p < 0.001). Six fractures suspected in radiographs were ruled out by CBCT. Treatment changes based on additional information from CBCT were recommended in 29 patients (31.5%). While agreement between advised therapy before CBCT and actual treatment was moderate (κ = 0.41 [95% confidence interval 0.35-0.47]; p < 0.001), agreement after CBCT was almost perfect (κ = 0.88 [0.83-0.93]; p < 0.001). Diagnostic confidence increased considerably for CBCT studies (p < 0.001). Median effective dose for CBCT was 4.3 μSv [3.3-5.3 μSv] compared to 0.2 μSv [0.1-0.2 μSv] for radiography.
Conclusions: CBCT provides advantages for the evaluation of acute small bone and joint trauma by detecting and excluding extremity fractures and fracture-related findings more reliably than radiographs. Additional findings induced therapy change in one third of patients, suggesting substantial clinical impact.
Key Points: • With cone-beam CT, extremity fractures and fracture-related findings can be detected and ruled out more reliably than with conventional radiography. • Additional diagnostic information provided by cone-beam CT scans has substantial impact on therapy in small bone and joint trauma. • For distal extremity injury assessment, one-stop-shop imaging without repositioning is feasible with the twin robotic x-ray system.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies