Tytuł pozycji:

Agreement between hematocrits measured with electrical conductivity on the i-STAT analyzer and manual packed cell volume measurements in 148 chelonian patients.

Tytuł:
Agreement between hematocrits measured with electrical conductivity on the i-STAT analyzer and manual packed cell volume measurements in 148 chelonian patients.
Autorzy:
Wong S; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Oklahoma State University, Stillwater, OK, USA.
Di Girolamo N; Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Oklahoma State University, Stillwater, OK, USA.; Tai Wai Small Animal & Exotic Hospital, Sha Tin, Hong Kong.
Źródło:
Veterinary clinical pathology [Vet Clin Pathol] 2021 Sep; Vol. 50 (3), pp. 342-347. Date of Electronic Publication: 2021 Sep 20.
Typ publikacji:
Journal Article
Język:
English
Imprint Name(s):
Publication: 1998- : Baton Rouge, LA : American Society for Veterinary Clinical Pathology
Original Publication: 1977-<1989> : Columbia, MO : Veterinary Practice Pub. Co.
MeSH Terms:
Point-of-Care Systems*
Turtles*
Animals ; Cell Size ; Electric Conductivity ; Hematocrit/veterinary
References:
Grosenbaugh DA, Gadawski KE, Muir WW. Evaluation of a portable clinical analyser in a veterinary hospital setting. J Am Vet Med Assoc. 1998;213:691-694.
Bingham D, Kendall J, Clancy M. The portable laboratory: an evaluation of the accuracy and reproducibility of i-STAT. Ann Clin Biochem. 1999;36:66-71.
Woolley A, Hickling K. Errors in measuring blood gases in the intensive care unit: effect of delay in estimation. J Crit Care. 2003;18:31-37.
Longanbach S, Miers M. Automated blood cell analysis. In: Keohane E, Smith L, Walenga J, eds. Rodak’s hematology: Clinical principles and applications (5th ed.). Elsevier Saunders; 2016:208-234.
Stockham SL, Scott MA. Fundamentals of veterinary clinical pathology. John Wiley & Sons; 2013.
Lewbart GA, Hirschfeld M, Denkinger J, et al. Blood gases, biochemistry, and hematology of Galapagos green turtles (Chelonia mydas). PLoS One. 2014;9(5):e96487.
Lewbart GA, Griffioen JA, Savo A, et al. Biochemistry and hematology parameters of the San Cristóbal Galápagos tortoise (Chelonoidis chathamensis). Con Phys. 2018;6:1-8.
Eshar D, Gancz AY, Avni-Magen N, King Rm Beaufrère H. Hematologic, plasma biochemistry and acid-base analysis of adult Negev desert tortoises (Testudo werneri) in Israel. J Zoo Wildl Med. 2014;45(5):979-983.
Muñoz-Pérez JP, Lewbart GA, Hirschfeld M, et al. Blood gases, biochemistry and haemotology of Galápagos hawksbill turtles (Eretmochelys imbricata). Con Phys. 2017;5:1-9.
Wolf KN, Harms CA, Beasley JF. Evaluation of five clinical chemistry analysers for use in health assessment in sea turtles. J Am Vet Med Assoc. 2008;233:470-475.
Di Girolamo N, Ferlizza E, Selleri P, Nardini G, Isani G. Evaluation of point-of-care analysers for blood gas and clinical chemistry in Hermann’s tortoise (Testudo hermanni). J Small Anim Pract. 2018;59:704-713.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet. 1986;1:307-310.
Colon VA, Di Girolamo N. Prognostic value of packed cell volume and blood glucose concentration in 954 client-owned chelonians. J Am Vet Med Assoc. 2020;257:1265-1272.
Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135-160.
Papadea C, Foster J, Grant S, et al. Evaluation of the i-STAT Portable Clinical Analyser for point-of-care blood testing in the intensive care units of a university children's hospital. Ann Clin Lab Sci. 2002;32:231-243.
Westgard JO, Hunt MR. Use and interpretation of common statistical tests in method-comparison studies. Clin Chem. 1973;19:49-57.
510(k) substantial equivalence determination decision summary. i-STAT Hematocrit test. https://www.accessdata.fda.gov/cdrh_docs/reviews/K163342.pdf. Accessed February 11, 2021.
Hopfer SM, Nadeau FL, Sundra M, Makowski GS. Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods. Ann Clin Lab Sci. 2004;34(1):75-82.
Contributed Indexing:
Keywords: agreement; chelonians; exotics; method comparison; reptiles; zoological
Entry Date(s):
Date Created: 20210921 Date Completed: 20210930 Latest Revision: 20210930
Update Code:
20240105
DOI:
10.1111/vcp.13014
PMID:
34546574
Czasopismo naukowe
Background: Point-of-care (POC) analyzers are increasingly used for diagnostic testing in exotic animals. A few studies with small sample sizes, evaluating agreement between hematocrit (Hct) and packed cell volume (PCV) measurements in chelonians, showed conflicting results.
Objective: We aimed to assess the agreement between Hct values measured with a POC analyzer and manual PCV measurements in venous samples from 148 chelonians.
Methods: All chelonians that underwent bloodwork for clinical reasons had Hct measured with an analyzer employing electrical conductivity (i-STAT, Abbott), PCVs measured using centrifugation, and total solids (TS) measured with refractometry. Bland-Altman plots were built to assess agreement between the measurements. Generalized linear mixed models were used to determine the effect of different variables on the difference between Hct and PCV.
Results: Of 192 chelonians sampled during this period, 148 had Hct, PCV, and TS results. Hct significantly underestimated PCV, with a mean difference between the techniques of -3.8% (95% CI: -4.5 to -3.2; LoA: -11.5 to 3.8). There was a negative correlation between the value of TS and the difference between the methods. Controlling for other factors, the magnitude of the disagreement was not affected by the sex of the chelonian but was smaller in red-eared sliders (Trachemys scripta elegans).
Conclusions: Hct values measured with electrical conductivity in chelonians significantly underestimated manual PCVs by about 4%. This difference should be considered whenever an Hct measured on a POC analyzer using electrical conductance is used in chelonian patient assessments.
(© 2021 American Society for Veterinary Clinical Pathology.)

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies