The purpose of this study is to compare a number of measured image quality parameters using processed and unprocessed or raw images in two full-field direct digital units and one computed radiography mammography system. This study shows that the difference between raw and processed image data is system specific. The results have shown that there are no significant differences between raw and processed data in the mean threshold contrast values using the contrast-detail mammography phantom in all the systems investigated; however, these results cannot be generalised to all available systems. Notable differences were noted in contrast-to-noise ratios and in other tests including: response function, modulation transfer function , noise equivalent quanta, normalised noise power spectra and detective quantum efficiency as specified in IEC 62220-1-2. Consequently, the authors strongly recommend the use of raw data for all image quality analyses in digital mammography. [ABSTRACT FROM AUTHOR]
Copyright of Radiation Protection Dosimetry is the property of Oxford University Press / USA and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)