Comparison of stone elimination capacity and drilling speed of endoscopic clearance lithotripsy devices.

Tytuł:
Comparison of stone elimination capacity and drilling speed of endoscopic clearance lithotripsy devices.
Autorzy:
Bader, Markus J.
Eisel, Maximilian
Strittmatter, Frank
Nagele, Udo
Stief, Christian G.
Pongratz, Thomas
Sroka, Ronald
Training and Research in Urological Surgery and Technology (T.R.U.S.T.)-Group
Temat:
ULTRASONIC equipment
LITHOTRIPSY
SPEED
LASER lithotripsy
Źródło:
World Journal of Urology; Feb2021, Vol. 39 Issue 2, p563-569, 7p
Czasopismo naukowe
Purpose: To investigate the fragmentation capacity, clearance time, and drilling speed of combined ultrasonic with impact dual-energy and single energy ultrasonic lithotripter devices. Methods: Stone fragmentation and clearance tests were performed under direct view in an underwater layered hemisphere by four different operators using artificial stones (n = 10/operator). Time for complete clearance was measured. Drilling tests were performed using an underwater setup, consisting of a mounting rack for fixing the lithotripter handpiece with the probe in vertical position and in contact with the stone phantom placed on one side of a balance for defined and constant contact application pressure equivalent to 450 g load. Time until complete perforation or in case of no perforation, the penetration depth after 60 s into the stone sample was recorded. Four devices, one single energy device (SED), one dual-energy dual probe (DEDP), two dual-energy single probe (DESP-1, DESP-2), with different parameters were tested. Results: Stone fragmentation and clearance speed were significantly faster for dual-energy device DESP-1 compared to all other devices (p < 0.001). Using DESP-1, the clearance time needed was 26.0 ± 5.0 s followed by DESP-2, SED and DEDP requiring 38.4 ± 5.8 s, 40.1 ± 6.3 s and 46.3 ± 11.6 s, respectively. Regarding the drilling speed, DESP-1 was faster compared to all other devices used (p < 0.05). While the drilling speed of DESP-1 was 0.69 ± 0.19 mm/s, compared to 0.49 ± 0.18 mm/s of DESP-2, 0.47 ± 0.09 mm/s of DEDP, and 0.19 ± 0.03 mm/s of SED. Conclusions: The dual-energy/single-probe device combining ultrasonic vibrations with electromechanical impact was significantly faster in fragmentation and clearing stone phantoms as well as in drilling speed compared to all other devices. [ABSTRACT FROM AUTHOR]
Copyright of World Journal of Urology is the property of Springer Nature and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies