Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Tytuł pozycji:

Distortion‐free, high‐isotropic‐resolution diffusion MRI with gSlider BUDA‐EPI and multicoil dynamic B0 shimming.

Tytuł:
Distortion‐free, high‐isotropic‐resolution diffusion MRI with gSlider BUDA‐EPI and multicoil dynamic B0 shimming.
Autorzy:
Liao, Congyu
Bilgic, Berkin
Tian, Qiyuan
Stockmann, Jason P.
Cao, Xiaozhi
Fan, Qiuyun
Iyer, Siddharth Srinivasan
Wang, Fuyixue
Ngamsombat, Chanon
Lo, Wei‐Ching
Manhard, Mary Kate
Huang, Susie Y.
Wald, Lawrence L.
Setsompop, Kawin
Temat:
DIFFUSION magnetic resonance imaging
HIGH resolution imaging
PHASE coding
Źródło:
Magnetic Resonance in Medicine; Aug2021, Vol. 86 Issue 2, p791-803, 13p
Czasopismo naukowe
Purpose: We combine SNR‐efficient acquisition and model‐based reconstruction strategies with newly available hardware instrumentation to achieve distortion‐free in vivo diffusion MRI of the brain at submillimeter‐isotropic resolution with high fidelity and sensitivity on a clinical 3T scanner. Methods: We propose blip‐up/down acquisition (BUDA) for multishot EPI using interleaved blip‐up/blip‐down phase encoding and incorporate B0 forward‐modeling into structured low‐rank reconstruction to enable distortion‐free and navigator‐free diffusion MRI. We further combine BUDA‐EPI with an SNR‐efficient simultaneous multislab acquisition (generalized slice‐dithered enhanced resolution ["gSlider"]), to achieve high‐isotropic‐resolution diffusion MRI. To validate gSlider BUDA‐EPI, whole‐brain diffusion data at 860‐μm and 780‐μm data sets were acquired. Finally, to improve the conditioning and minimize noise penalty in BUDA reconstruction at very high resolutions where B0 inhomogeneity can have a detrimental effect, the level of B0 inhomogeneity was reduced by incorporating slab‐by‐slab dynamic shimming with a 32‐channel AC/DC coil into the acquisition. Whole‐brain 600‐μm diffusion data were then acquired with this combined approach of gSlider BUDA‐EPI with dynamic shimming. Results: The results of 860‐μm and 780‐μm datasets show high geometry fidelity with gSlider BUDA‐EPI. With dynamic shimming, the BUDA reconstruction's noise penalty was further alleviated. This enables whole‐brain 600‐μm isotropic resolution diffusion imaging with high image quality. Conclusions: The gSlider BUDA‐EPI method enables high‐quality, distortion‐free diffusion imaging across the whole brain at submillimeter resolution, where the use of multicoil dynamic B0 shimming further improves reconstruction performance, which can be particularly useful at very high resolutions. [ABSTRACT FROM AUTHOR]
Copyright of Magnetic Resonance in Medicine is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies