A comparison of methods for estimating the line spread function of a CT imaging system

Tytuł:
A comparison of methods for estimating the line spread function of a CT imaging system
Autorzy:
Abboud, Samir
Lee, Kristina
Vinehout, Kaleb
Paquerault, Sophie
Kyprianou, Iacovos S.
Źródło:
Proceedings of SPIE; March 2011, Vol. 7961 Issue: 1 p79615V-79615V-10, 7881896p
Periodyk
A quantitative description of the deterministic properties of a CT system is necessary when evaluating image quality. The most common such metric is the modulation transfer function (MTF), usually calculated from a line spread function (LSF) or point spread function (PSF). Currently, there exist many test objects used to measure the LSF or PSF. In this paper we report on a comparison of these measures using: a thin foil slit test object, a teflon cube edge test object and a novel "negative" cube test object. Images were acquired using a custom-built bench-top flat-panel-based cone-beam CT scanner and a cylindrical water-filled PMMA phantom with the test objects embedded in the middle. From the 3-dimensional reconstructed volumes, we estimated the LSF either directly or as estimated from the edge spread function. From these, a modulation transfer function can be estimated, and the frequency dependent image transfer of each object can be reported.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies