Informacja

Drogi użytkowniku, aplikacja do prawidłowego działania wymaga obsługi JavaScript. Proszę włącz obsługę JavaScript w Twojej przeglądarce.

Tytuł pozycji:

Successful retrieval of a retained J guidewire from the right atrium and aorta arch using a self-invented snare device

Tytuł:
Successful retrieval of a retained J guidewire from the right atrium and aorta arch using a self-invented snare device
Autorzy:
Chuan Chen, MD,PhD
Robin Bhattarai, MS,PhD
Yang Yang, MD
Cong Ling, MD,PhD
Lun Luo, MD
Hui Wang, MD,PhD
Temat:
retained j guidewire
retrieval
self invented snare device
aneurysm embolization
thromboembolic event
Surgery
RD1-811
Neurosciences. Biological psychiatry. Neuropsychiatry
RC321-571
Źródło:
Nepal Journal of Neuroscience, Vol 18, Iss 2, Pp 67-70 (2021)
Wydawca:
NESON, 2021.
Rok publikacji:
2021
Kolekcja:
LCC:Surgery
LCC:Neurosciences. Biological psychiatry. Neuropsychiatry
Typ dokumentu:
article
Opis pliku:
electronic resource
Język:
English
ISSN:
1813-1948
1813-1956
12853259
Relacje:
https://www.nepjol.info/index.php/NJN/article/view/35728; https://doaj.org/toc/1813-1948; https://doaj.org/toc/1813-1956
DOI:
10.3126/njn.v18i2.35728
Dostęp URL:
https://doaj.org/article/b17d8e24a0234b128532596e66f0f326  Link otwiera się w nowym oknie
Numer akcesji:
edsdoj.b17d8e24a0234b128532596e66f0f326
Czasopismo naukowe
Guidewire retention during an interventional treatment is a very rare complication that can lead to possible serious complications, such as thrombosis, embolization, sepsis, and perforation. Hence, a retained guidewire must be retrieved as soon as possible. Here, we provide the first report of a case of accidental retention of a 0.035” J guidewire in the right atrium and aortic arch during basilar artery aneurysm embolization. We found that due to the heavier weight of the J guidewire and its hydrophilic-coated layer, it could not be removed using previously reported methods for micro-guidewire retention. Therefore, we constructed a self-invented snare device consisting of a 0.014” micro-guidewire and a 4-Fr MP-A1 catheter and successfully retrieved the J guidewire, thus avoiding open surgery and its associated serious complications. According to our experience, this novel technique described here is quite easy to perform, safe and effective and can avoid serious complications and prevent the need for open surgery.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies